SEARCH TELATOMIC

PO Box 924, Jackson, MI   49204       1 (800) 622-2866

HOME


CWS 12-50 SPECTROMETER SPECIFICATIONS

Mode NMR (1H, 19F), ESR
Operational Frequency NMR 1H-13 MHz, NMR 19F-12.2 MHz, ESR-50 MHz
Frequency Stability ≤ 1 PPM/ 1h
Electromagnet  
    - magnetic field 300 mT
    - maximum current 0.7A
    - coil 2,000 turns
    - gap 10.5 mm
    - pole diameter 50 mm
    - homogeneity ≤ 10-4/sample volume
    - field stability ≤ 10 μT/1 hr
Helmholtz Coils  
    - magnetic field 195 μT
    - gap 15 mm
    - coils diameter 70 mm
    - homogeneity 10-5/sample volume
Modulation Field  
    - frequency 38 Hz
    - amplitude 0.1-20 μT
Sweep of Magnetic Field (NMR and ESR)  
    - range 0.5 mT - 10.0 mT
    - time 0.5 min - 30 min
Sweep of Frequency (only NMR)  
    - range 20 Hz - 400 kHz
    - time 0.5 min - 30 min
RF Probehead  
    - solenoid coil dimensions ID=5.8 mm; L=10 mm
    - mode Automatically tuned for NMR or ESR
Receiver  
    - gain 0-48 dB (2 dB step)
    - detection phase-sensitive
    - phase adjustment 0-360°, step 1.5°
    - signal filter sweep controlled
    - DC offset converter automatic
A/D Converter  
    - resolution 10 bit
    - number of samples Min 512 per sweep
Weight and dimensions WxDxH  
    - electronic unit 3.5 kg, 350x135x85 mm
    - probehead 0.4 kg, 35x210x70 mm
    - Helmholtz coils (ESR) 0.5 kg, 50x80x110 mm
    - electromagnet (NMR) 10.5 kg, 175x100x160 mm
Power Consumption 110V/220 V; 50/60 Hz; 40 W
Communication Port two way RS 232C
Computer Required IBM PC, min 750 MHz, VGA color or compatible
Software MS Windows operated


SAMPLES

NMR Sample Nuclei
  H2O + CuSO4 1H
  Rubber 1H
  Acrylic 1H
  Delrin 1H
  HBF4 1H and 19F
  Teflon 19F
 
ESR Sample (free radicals)
  TCNQ
DPPH
TMPM


SPECTROMETER

Operation:

Recording of NMR and ESR spectra

From recorded spectra one can determine:

1. Second moment
2. Line width
3. g factor
4. Line shape
5. Earth magnetic field magnitude
6. Accurate value of the magnetic field in electromagnet and Helmholtz coil


COMPONENT DETAIL










Need technical information or to place an order?
TEL-Atomic     1-800-622-2866
Teltron
Products

Tel-X-Ometer X-Ray
  & Accessories
Electron Tubes

Spectroscope

Thermodynamics
Radiometer
Leslie's Cube
Stefan-Boltzmann
Nuclear
Systems

Multichannel Analyzer
Counters
Basic Systems
Advanced System
Isotope Generator

Spin Resonance
Systems

Pulsed NMR
Advanced NMR/ESR
Other Products
Kater Pendulum
Photogate
Air Tracks
Oscilloscopes
Instrumentation
  Sweep Generator
    with Counter
  Multimeters
Superconductivity
Photoelectric Effect
Cloud Chamber
Planck's Constant
Cavendish Balance
Rutherford Scattering Apparatus
Speed of Light Apparatus
Magnetic Bubble Apparatus
Millikan Apparatus
Smart Magnetic Sensor
Power Supplies

Warranty, Freight & Terms     Contact TEL-Atomic
sales@telatomic.com